Acute and Long-term Results After Contemporary Subcutaneous Implantable Cardioverter-defibrillator Implantation: A Single-center Experience

Miguel A. Arias a,, Marta Pachón a, Finn Akerström a, Alberto Puchol a, Cristina Martín-Sierra a, Luis Rodríguez-Padial a

a Unidad de Arritmias y Electrofisiología Cardiaca, Servicio de Cardiología, Hospital Virgen de la Salud, Toledo, Spain

Refers to

The Subcutaneous ICD. Ready to Conquer Everyone's Heart?
Lucas V.A. Boersma
Rev Esp Cardiol. 2018;71:886-7
Full text - PDF

Keywords

Subcutaneous defibrillator. Cardiac arrest. Ventricular arrhythmias.

Abstract

Introduction and objectives

The subcutaneous implantable cardioverter-defibrillator (S-ICD) has emerged as an alternative to the transvenous defibrillator. The incidence of complications is similar, with inappropriate shocks (IS) being more frequent than those occurring with contemporary programming of transvenous defibrillators. Several improvements have been implemented after the S-ICD was approved for use in Europe in 2009. This study reports the results of S-ICD use in a single center, whose experience began late, at the end of 2013.

Methods

Prospective observational study including consecutive patients with defibrillator indication and no indication for either permanent pacing or cardiac resynchronization who underwent S-ICD implantation. Implant data and long-term follow-up were analyzed.

Results

An S-ICD was implanted in 50 patients who were deemed suitable after electrocardiographic screening. The mean age was 46.9 ± 15 (range, 15-78) years and 72% were male. Thirty eight percent had left ventricular ejection fraction ≤ 35%. The most frequent heart disease was ischemic heart disease (34%), followed by hypertrophic cardiomyopathy (18%). The intermuscular technique was used, with 3 incisions in 10% and 2 incisions in the remaining 90%. Ventricular fibrillation was induced in 49 patients, with 100% effectiveness in their conversion. After a mean follow-up of 18.1 (range, 2.3-44.8) months, there were no late complications requiring surgical revision, the rate of IS was 0%, and 1 patient (2%) experienced appropriate shocks.

Conclusions

Improvements in technology, implant technique and device programming, along with appropriate patient selection, have led to outstanding acute and long-term results, especially regarding the absence of both IS and complications requiring surgical revision.

Article

You can access the full text of this article:

Members of the Spanish Society of Cardiology


Subscribers of Revista Española de Cardiología

You may purchase this article for 30.00 €

Buy Now
Subscribers of ScienceDirect

1885-5857/© 2018 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved

Cookies
x
To improve our services and products, we use cookies (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.
Cookies policy
x
To improve our services and products, we use cookies (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.