Rev Esp Cardiol. 2018;71:79-85 - Vol. 71 Num.02 DOI: 10.1016/j.rec.2017.04.011

The Girona Territori Cardioprotegit Project: Performance Evaluation of Public Defibrillators

Pablo Loma-Osorio a,b,, Maria Nuñez a, Jaime Aboal a, Daniel Bosch a, Pau Batlle c, Ester Ruiz de Morales c, Rafael Ramos b,d, Josep Brugada e, Hisao Onaga f, Alex Morales c, Josep Olivet g, Ramon Brugada a,b,h

a Servicio de Cardiología, Hospital Universitari Dr. Josep Trueta, Girona, Spain
b Departamento de Ciencias Médicas, Facultad de Medicina, Universidad de Girona, Girona, Spain
c Departamento de Salud, Diputación de Girona (Dipsalut), Girona, Spain
d Instituto de Investigación en Atención Primaria (IDIAP Jordi Gol), Girona, Spain
e Instituto del Tórax, Hospital Clínic i Provincial de Barcelona, Barcelona, Spain
f Servicio de Emergencias Médicas (SEM), Región Sanitaria de Girona, Girona, Spain
g Grupo de Investigación, Salud y Atención Sanitaria, Universidad de Girona, Girona, Spain
h Centro de Genética Cardiovascular, Instituto de investigación biomédica de Girona (IDIBGI), Girona, Spain

Refers to

Keywords

Public defibrillation. Cardiopulmonary arrest. Cardiopulmonary resuscitation. Ventricular fibrillation. Asystole.

Abstract

Introduction and objectives

In recent years, public access defibrillation programs have exponentially increased the availability of automatic external defibrillators (AED) in public spaces but there are no data on their performance in our setting. We conducted a descriptive analysis of the performance of AED since the launch of a public defibrillation program in our region.

Methods

A retrospective analysis was conducted of electrocardiographic tracings and the performance of AED in a public defibrillation program from June 2011 to June 2015 in the province of Girona, Spain.

Results

There were 231 AED activations. Full information was available on 188 activations, of which 82% corresponded to mobile devices and 18% to permanent devices. Asystole was the most prevalent rhythm (42%), while ventricular fibrillation accounted for 23%. The specificity of the device in identifying a shockable rhythm was 100%, but there were 8 false negatives (sensitivity 83%). There were 47 shockable rhythms, with a spontaneous circulation recovery rate of 49% (23 cases). There were no accidents related to the use of the device.

Conclusions

Nearly half of the recorded rhythms were asystole. The AED analyzed showed excellent safety and specificity, with moderate sensitivity. Half the patients with a shockable rhythm were successfully treated by the AED.