Rev Esp Cardiol. 2017;70:706-12 - Vol. 70 Num.09 DOI: 10.1016/j.rec.2017.03.008

Atrial Fibrillation Ablation Guided by a Novel Nonfluoroscopic Navigation System

Gabriel Ballesteros a, Pablo Ramos a, Renzo Neglia a, Diego Menéndez a, Ignacio García-Bolao a,

a Departamento de Cardiología y Cirugía Cardiaca, Clínica Universidad de Navarra, Pamplona, Navarra, Spain

Refers to

Keywords

Atrial fibrillation ablation. Pulmonary vein isolation. High-density electroanatomic mapping. Rhythmia Mapping System. Orion catheter.

Abstract

Introduction and objectives

Rhythmia is a new nonfluoroscopic navigation system that is able to create high-density electroanatomic maps. The aim of this study was to describe the acute outcomes of atrial fibrillation (AF) ablation guided by this system, to analyze the volume provided by its electroanatomic map, and to describe its ability to locate pulmonary vein (PV) reconnection gaps in redo procedures.

Methods

This observational study included 62 patients who underwent AF ablation with Rhythmia compared with a retrospective cohort who underwent AF ablation with a conventional nonfluoroscopic navigation system (Ensite Velocity).

Results

The number of surface electrograms per map was significantly higher in Rhythmia procedures (12 125 ± 2826 vs 133 ± 21 with Velocity; P < .001), with no significant differences in the total procedure time. The Orion catheter was placed for mapping in 99.5% of PV (95.61% in the control group with a conventional circular mapping catheter; P = .04). There were no significant differences in the percentage of PV isolation between the 2 groups. In redo procedures, an ablation gap could be identified on the activation map in 67% of the reconnected PV (40% in the control group; P = .042). The measured left atrial volume was lower than that calculated by computed tomography (109.3 v 15.2 and 129.9 ± 13.2 mL, respectively; P < .001). There were no significant differences in the number of complications.

Conclusions

The Rhythmia system is effective for AF ablation procedures, with procedure times and safety profiles similar to conventional nonfluoroscopic navigation systems. In redo procedures, it appears to be more effective in identifying reconnected PV conduction gaps.